Page 10 - Voltech Instruments PM3300 Universal Power Analyser, designed and produced by Connell Marketing Associated, UK
P. 10

Wiring and Channel Select

                                                           WIRING CONFIGURATION
                                                                          shunt

                                                                   1 phase 2 wire
                                                                   1 phase 3 wire
                                                                   3 phase 3 wire
                                                                   3 phase 4 wire
                                                                    channel 3 only
                                                                    channel 2 only
                                                                   independent ch3
                                                               independent channels

    The PM3300 can be configured to analyze almost any electrical system. Its versatility
    is achieved because all the power inputs are ac + dc coupled and isolated from each
    other and from ground.

    The inputs can therefore be connected as required by the user via the main menu.

    Frequency and auto ranging are normally set for all channels based on the inputs to
    channel 1.

    The independent wiring modes allow inputs of differing magnitude or frequency to be
    monitored simultaneously. This is very useful for frequency converters and DC bus
    measurements. These modes are used for load efficiency measurements (see
    page 15).

    shunt     This selects internal or external shunt mode:

              Internal is selected for direct measurements of ac and dc
              currents up to 30Arms, and uses signals from precision internal
              shunts. This mode can also be used with AC current
              transformers.

              External is selected when taking current measurements using
              external high current shunts, or other devices providing a
              voltage signal proportional to current.

    1 phase 2 wire Select for single phase loads, 2 wire distribution systems and
                               rectifier outputs.

    1 phase 3 wire Select for three wire single phase distribution systems.

    3 phase 3 wire Select for three phase loads without neutral connection (2
                               wattmeter method).

    3 phase 4 wire Select for three phase loads with or without neutral connection
                               (3 wattmeter method).

    independent ch3 Select, for example, for 3 phase input on channels 1 and 2 (Two
                                wattmeter) and DC bus on Channel 3.

    independent channels Select for completely independent channel operation, such as
                                   for input, DC bus and output of a frequency converter.

              CHANNEL SELECT

              These keys select the channel measurements to
              be displayed.

           Σ  SUMMATION

              Displays the total (summation value for the multiphase system).

                    NEUTRAL
           N Provides, analysis of the neutral currents by computation from

                      the individual phase currents.

                      Arms, Apeak and Acf can be displayed. The THD and harmonic
                      analysis can be performed.