Harmonic Analysis
CH1
AH03 = 4 . 725 A - 188 . 4°
> VH03 = 289 . 6mV - 252 . 2°
A thd = 89 . 09 %
V thd = 1 . 170 %
Freq = 61 . 03 Hz
PF = 0 . 746
The PM3300 provides comprehensive harmonic analysis capability.
It will measure the harmonic content of the voltage, current and real power on each
phase, for all components up to the 99th harmonic. Power harmonics are available
via the internal printer.
The Harmonic Barchart display mode provides a convenient way of viewing the
spectrum of distortion components (see page 5).
Analysis of neutral currents and DC components in the waveforms can also be
performed.
Magnitude of harmonic as absolute Phase angle of harmonic referenced
value or % of fundamental to voltage on CH1
CH1
AH03 = 4 . 725 A - 188 . 4°
> VH03 = 289 . 6mV - 252 . 2°
A thd = 89 . 09 %
V thd = 1 . 170 %
Freq = 61 . 03 Hz
PF = 0 . 746
THD expressed as % of fundamental
or rms value of waveform
HARM HARMONICS
THD This key provides a display menu to select VOLTAGE or CURRENT
FREQ Harmonic and the Harmonic number to be displayed. The harmonics
FUND can then be scrolled through using the ENTER key. Harmonics can be
output to the printer or PC.
TOTAL HARMONIC DISTORTION
This key provides display menu to select VOLTAGE or CURRENT
total harmonic distortion.
FREQUENCY
Displays the frequency of the AC supply. This is the fundamental
frequency for harmonic analysis.
FUNDAMENTAL
Displays the fundamental component of all relevant displayed
measurement functions (see example page 5).