Set Up

The set up keys control the configuration of the analog circuitry of the PM3000A. This
enables the PM3000A to be used in a wide variety of specialized applications.

 FREQ     FREQ SOURCE.
SOURCE
          The measurements made by the PM3000A are synchronized to the ac
     INT  supply frequency.
     EXT
          INT mode synchronizes measurements to input voltage or current,
          over the frequency range 5Hz to 1MHz. The user can choose voltage
          or current for frequency measurement, which is helpful in applications
          such as motor drives, where one input may be 'chopped'.

          EXT mode synchronizes measurements to a signal applied to a rear
          panel connector. It is used when both voltage and current are
          'chopped', or for frequencies down to 0.1Hz.

VOLTS     VOLTS AMPS.
AMPS
          RMS This mode is normally selected whereby the PM3000A displays
    RMS   voltage and current measurements as true rms.
    MEAN
          MEAN Voltage and current measurements will be displayed in terms
          of their rectified mean value. This is useful in electrochemical
          applications and for calibrating mean sensing instruments.

AVG       AVG.

  AUTO    The PM3000A will average a number of measurements to display a
  N       steady reading despite fluctuations at the inputs.

          AUTO In this mode the instrument will average 16 measurements;
          any substantial change in input conditions will automatically cancel
          averaging to give a fast reponse.

          N In this mode the user can force the PM3000A to average up to 64
          measurements in order to provide a steady reading even when large
          changes in input conditions are occuring.

COUP      COUPLING
- LING
          AC + DC is normally selected. The PM3000 will compute results due
          to all components in the input voltage and current waveforms, including

AC+DC the dc and low frequency components.

            AC configures the instrument to ac couple the input signals. This is
AC useful for examining ripple voltage on rectifier outputs and for the

          calibration of other instruments that are ac coupled.

SHUNT     SHUNT

    INT   INT mode is selected for direct measurements of ac and dc currents
    EXT   up to 30A rms, and uses signals from precision internal shunts in the
          PM3000A. This mode is also used with AC current transformers.

          EXT mode is used when taking current measurements with external
          high current shunts, DC current transformers, or other devices providing
          a voltage signal that is proportional to current.